Description
Quickem POCT Vet is a fully automated chemistry analyzer designed for the on-site testing. It can provide faster, easier, more accurate medical diagnosis using 0.1cc of whole blood, serum or plasma. Utilizing cloud computing, SD1 can help the operators to make medical decisions remotely. Built-in centrifuge, QR code , real-time quality control. The test results will be automatically printed after 12 minutes.
Small sample required
Only 100ul samples are needed for a test, compared to a traditional chemistry analyzer, the sample volume is only 1/10.
Easy to use
No pre-processing for sample. It is extremely easy to use does not require dedicated staff and allows anyone could operate it.
Automatic/ Manual Barcode Identity
Read the Reagent Discs Automatically, Confirm the expiry date of the reagent automatically
Also Manual Barcode Identity – load Patient information via LIS automatically
Accuracy
Adopt the photoelectric colorimetric principle with a built-in real-time quality control system to ensure the accuracy of the analysis.
Quick result
It is real-time test and you could read the test report in 12 minutes, which greatly shortens the waiting time for patients.
Automated and smart
It is fully automated with features of online fault diagnosis, upgrading, to help you know the operation of the instrument in real time.
Compact and portable
The analyzer is small and lightweight which will save your laboratory space and allow on-site patient testing.
Minimal maintenance
There are no easy-wearing parts such as liquid circuit, pump valve, etc., and the single reagent tray is used as a complete set, which is free of cleaning and maintenance.
Auto sleep and one-click awake, one-click fault clearing, Status monitoring and Alarm
Easy data management
Memory capacity up to 500,000 patient records, through RS-232 port, it can be connected to the LIS laboratory information system to facilitate customer data management.
SPECIFICATIONS:
- Sample Type
- Anti-coagulation whole blood, serum, plasma
- Sample Volume
- 90-120ul
- Bar Code
- QR Code
- Testing Time
- 12 minutes/sample
- Testing Principle
- Absorption spectroscopy, transmission turbidimetry
- Testing Method
- Endpoint, kinetic, fixed time, turbidimetry, etc
- Temperature
- 37±0.3℃
- Resolution
- Abs
- Cross Contamination
- 0
- QC & Calibrate
- Intelligent QC
- Working Condition
- Temperature:10-30℃ Humidity: 30%-70%
- Light Source
- 12V/20W, halogen tungsten lamp with lifespan over 2500 hours
- Power Supply
- Input: AC 220V, 50Hz
- Power Rating
- 120VA
- Optical System
- After the filter spectral, 8 wavelength synchronous detection: 340, 405, 450, 505, 546, 600, 630, 850nm
- Display
- Android 7.0 inch 800*480, multi-point capacitive touch screen
- Storage
- 500,000 results
- Printer
- Built-in thermal printer
- Connectors
- 4 USB ports, 1 LAN port
- Net Weight
- 6 Kg
- Dimension (L*W*H)
- 200mm*252mm*299mm
Profiles | Parameters ( * are Calculated ones) |
16 Comprehensive Diagnostic Parameters | ALB, TP, TBIL, ALT, AST, GLO, A/G, UREA, CREA, CK, AMY, GLU, TC, PHOS, Ca+, U/C |
11 Liver function Parameters | ALB, TP, TBIL, ALP, ALT, AST, GGT, TBA, GLO, A/G, TC |
10 Kidney function Parameters | ALB, UREA, CREA, UA, GLU, TCO2, PHOS, Ca+, K+, U/C |
10 Pre-Operation test Parameters | TP, ALP, AST, ALT, UREA, CREA, CK, LDH, GLU, U/C |
8 Kidney function Parameters | ALB, UREA, CREA, TCO2, PHOS, Ca+, K+, U/C |
6 Canine Inflammation Parameters | |
9 Feline Inflammation Parameters | |
Diabetes Test Parameters | |
Pancreatitis Test Parameters | |
Blood Gas Parameters | |
Electrolyte Parameters |